Yazarlar : Portakal O, Tavil B, Kuşkonmaz B, Aytaç S, Hasçelik G.
Yayın : J Clin Lab Anal.
Yayın Yılı : 2011
Pubmed Linki : http://www.ncbi.nlm.nih.gov/pubmed/21437995
Konu : Diğer
Literatür İçeriği :
This study was aimed to evaluate the analytical performance of an automated image analysis system (a pilot model of Diff Master(™) Octavia) for the preclassification of leucocytes in children with hematological disease. Manual microscopy performed by pediatric hematologists was used as the reference method. Five mature cell class and blasts were evaluated. Diff Master Octavia correctly preclassified 87.4% of all leucocytes with a high reproducibility. The overall accuracy was found to be 93.0%. Clinical sensitivity was 97.7% and specificity was 76.0%. The average time per slide for Diff Master(™) Octavia was 2.3 min lower than that of manual method. Our results indicated that the Diff Master(™) Octavia can detect and preclassify leucocytes accurately; therefore, it can be used as an efficient and fast method in pediatric hematology routine.
© 2011 Wiley-Liss, Inc.